OxyTechw2 RDO Pro-X Dissolved Oxygen Sensor


Quantity: Out Of Stock

Category: WaterWatch2 Monitors

Quick Overview:

Optical Sensor for Dissolved Oxygen Monitoring


Physical
Dimensions (hxwxd) Ø47mm, 203 mm long
Weight 0.8 kg
Protection Class IP68
Enclosure Material Body - ABS, Titanium Temperature Probe with Polyurethane coated cable

Cap - ABS, Polycarbonate/Polymethyl/methylacrylate
Cable Entries Integral Cable Gland, Black Acetal, 316 Stainless Steel
Cable Length 10 metres standard, 100 metres maximum
Interface to Monitor  
Power Supply 12 VDC from Monitor
Sensor Output ModTechw², RS485
Environmental Data  
Operating/Storage Temperature 0 to 50ºC
Pressure Rating 210 mWC
Flow No flow necessary
Measurement Characteristics
Principal of Operation Blue/Red Luminescence Quenching
Light Source Blue LED
Accuracy +/- 0.1 mg/l on 0-8 mg/l range
+/- 0.2 mg/l on 8-20 mg/l range
+/- 10% reading 20 to 50 mg/l
Resolution 0.01 mg/l
Display Resolution see 7300w² Monitor
Response Time T90 <45 seconds, T95 <60 seocnds @ 25°C
Flowrate 1 cm/second
Units mg/l, ppm, %Sat
Range 0-50 mg/l (ppm)
0-200 %Sat
Temperature Measurement Thermistor
Temperature Accuracy ± 0.1°C
Temperature Resolution 0.01°C

Tag: #226770  #226665  #226661  #223980  #226668 

Share: Messenger


Optical Sensor for Dissolved Oxygen Monitoring


Physical
Dimensions (hxwxd) Ø47mm, 203 mm long
Weight 0.8 kg
Protection Class IP68
Enclosure Material Body - ABS, Titanium Temperature Probe with Polyurethane coated cable

Cap - ABS, Polycarbonate/Polymethyl/methylacrylate
Cable Entries Integral Cable Gland, Black Acetal, 316 Stainless Steel
Cable Length 10 metres standard, 100 metres maximum
Interface to Monitor  
Power Supply 12 VDC from Monitor
Sensor Output ModTechw², RS485
Environmental Data  
Operating/Storage Temperature 0 to 50ºC
Pressure Rating 210 mWC
Flow No flow necessary
Measurement Characteristics
Principal of Operation Blue/Red Luminescence Quenching
Light Source Blue LED
Accuracy +/- 0.1 mg/l on 0-8 mg/l range
+/- 0.2 mg/l on 8-20 mg/l range
+/- 10% reading 20 to 50 mg/l
Resolution 0.01 mg/l
Display Resolution see 7300w² Monitor
Response Time T90 <45 seconds, T95 <60 seocnds @ 25°C
Flowrate 1 cm/second
Units mg/l, ppm, %Sat
Range 0-50 mg/l (ppm)
0-200 %Sat
Temperature Measurement Thermistor
Temperature Accuracy ± 0.1°C
Temperature Resolution 0.01°C


Sorry, Currently No Video Uploaded.

OxyTechw2 RDO Pro-X Dissolved Oxygen Sensor


Please Leave Your Product Enquiry Here. We will Get Back To You Soon.

Product Specifications

  • Product SKU : -
  • Weight (GRAM) : -
  • Dimension : -