Orion STAR T930 Ion Titrator


Quantity: Out Of Stock

Category: Star T900 Series Potentiometric Titrators

Quick Overview:

Use the Orion Star T930 ion titrator for dedicated ion titrations including salt in foods, chloride in drinking and wastewater, ammonia and Total Kjeldahl Nitrogen (TKN) of wastewater, surfactants in consumer products and total hardness of water. Titration techniques include equivalence point titrations and preset mV endpoint titrations plus multiple known addition (MKA) mode. Using MKA mode, the titrator performs an automated calibration and calculation of sample concentration by adding aliquots of standard to sample, removing the need for a separate calibration and minimizing matrix effects.

Options:
START9300: 20mL burette, stirrer probe, dispenser probe, standard tubing kit, 1L plastic bottle, GL38 bottle cap with two ports and drying tube, computer cable, literature on USB drive, 110-240V power supplies

START9301: 9780SC silver billet electrode, 91CBNC electrode cable, 20mL burette, stirrer probe, dispenser probe, standard tubing kit, 1L plastic bottle, GL38 bottle cap with two ports and drying tube, computer cable, literature on USB drive, 110-240V power supplies

START9303: 8611BNWP ROSS sodium electrode, 20mL burette, stirrer probe, dispenser probe, standard tubing kit, 1L plastic bottle, GL38 bottle cap with two ports and drying tube, pint bottle cap, computer cable, literature on USB drive, 110-240V power supplies

START9304: 9512HPBNWP high performance ammonia electrode, 20mL burette, stirrer probe, dispenser probe, standard tubing kit, 1L plastic bottle, GL38 bottle cap with two ports and drying tube, pint bottle cap, computer cable, literature on USB drive, 110-240V power supplies

START9305: 9342BN surfactant electrode, 900200 reference electrode, 20mL burette, stirrer probe, dispenser probe, standard tubing kit, 1L plastic bottle, GL38 bottle cap with two ports and drying tube, computer cable, literature on USB drive, 110-240V power supplies

START9306: 9720BNWP calcium electrode, 20mL burette, stirrer probe, dispenser probe, standard tubing kit, 1L plastic bottle, GL38 bottle cap with two ports and drying tube, computer cable, literature on USB drive, 110-240V power supplies
Tag: #START9301  #START9303  #START9304  #START9305  #START9306 

Share: Messenger


Model Orion Star T930
Type Ion titrator
Description Orion Star T930 ion titrator total hardness kit
Includes 9720BNWP calcium electrode, 20mL burette, stirrer probe, dispenser probe, standard tubing kit, 1L plastic bottle, GL38 bottle cap with two ports and drying tube, computer cable, literature on USB drive, 110-240V power supplies
Applications Ion concentration titrations including calcium hardness of drinking water and wastewater and total hardness of drinking water and wastewater
Functions Available Sample titration, titrant standardization, calibration and direct measure
Advanced ISE Functions Multiple known addition (MKA); automated calibration and calculation of sample concentration by adding aliquots of standard to sample using the burette
Titration Technique Equivalence point or preset endpoint
Equivalence Points 1 to 2
Preset Endpoints 1 to 3
Titration Types Direct or back titration
Blank Options Fixed value or variable value using titration
Cycles Per Titration Run 1 to 5 cycles, option to exclude cycle results from average results and RSD calculations
Titrant Determination Standardization titration or manual concentration input
Titration Process Control Routine, quick, careful or user defined
Precision 0.5% RSD, dependent on environment and handling conditions
Titration Setup Wizard Yes, using Start a New Titration workflow
Burette Sizes 10 mL, 20 mL (included) or 50 mL
Burette Functions Automatic rinse cycles for flushing and discrete dispense volume with continuous option; Exceeds accuracy requirements of ISO8655-3
Resolution Burette drive offers advanced micro-step technology that provides 25,600 micro-steps per motor revolution for smooth and precise burette positioning (over 2 million micro-steps over the full burette stroke range)
Stirrer Probe Speeds 5 user selectable speeds (250 to 3700 RPM)
Methods Up to 10 methods with optional password protection; import/export methods via USB flash drive
Data Points 100 data sets each: sample titration, titrant standardization, calibration and direct measure
Data Management Details Data log export as CSV or PDF file, short or long format
Sample ID Auto-incremental, manual or off
Time and Date Yes, with nonvolatile battery backup
Memory Non-volatile memory preserves data logs and settings
Updateable Firmware Yes, using USB flash drive
Alarm Output Notifications for Titration Cycle Complete, Maximum Titrant Volume, Data Log Full, Calibration Due, Maintenance Reminder
Range (ISE) 0.0001 to 19990
Resolution (ISE) 0.0001 minimum, 1 to 4 significant digits (user selectable)
Accuracy (ISE) ±0.2 mV or ±0.05% of reading, whichever is greater
mV Range -2000.0 to +2000.0 mV
Resolution (mV) 0.1 mV
Accuracy (mV) ±0.2 mV
Range (Temperature) -5.0 to 100.0 °C; 23.0 to 212 °F
Resolution (Temperature) 0.1 °C
Accuracy (Temperature) ±0.2 °C
Calibration Ion concentration (ISE) and temperature
Calibration Points (ISE) 1 to 5 points
Temperature Probe Calibration Optional 1 point offset calibration
Temperature Selection Manual or automatic
Display Type 5.7” color capacitive touchscreen, 640 x 480 resolution
Backlight Option Yes, adjustable
Languages Chinese, English, French, German, Italian, Japanese, Korean, Portuguese and Spanish
Material ABS plastic
Channels 1
Probe Type Ion selective electrode (ISE), reference half-cell electrode, ATC temperature probe, stirrer probe
Inputs BNC (ion selective electrode), pin-tip (reference electrode), 8 pin MiniDIN (ATC temperature probe), stirrer input (stirrer probe)
Outputs 3 USB ports for printer, USB flash drive and computer software connections
Certifications/Compliance CE, TUV 3-in-1, FCC, EN/EIC61326-1, IEC 61010
IP Rating IP-51
Electrical Requirements 100-240V, 50/60 Hz
Dimensions (L x W x H) 10” x 16” x 14”; 25.4 cm x 40.6 cm x 35.6 cm
Temperature (English) Operating 41 to 104 °F
Temperature (Metric) Operating 5 to 40 °C
Relative Humidity Range 5 to 85%, non-condensing
Warranty 1 year replacement warranty
Unit Size -


Automated vs. Manual Titration

Automated Titrator Interface Start-Up

Orion Star T900 Series - Titrator Set up Guide

Thermo Scientific Orion Star T900 Series Titrators

Quick Start Guide Instructions: Orion Star T900 Titrator

Total Alkalinity in Water by Automatic Titration

Orion STAR T930 Ion Titrator


Please Leave Your Product Enquiry Here. We will Get Back To You Soon.

Product Specifications

  • Product SKU : -
  • Weight (GRAM) : -
  • Dimension : -